📧 info@texaslemonlaw.com 📞 +1 866 777 5045
Free Consultation Available

HORTON AMBULANCE 503

View complaints and recalls by year, or see all recent reports below.

HORTON AMBULANCE 503 Summary

4

Model Years

0

Total Complaints

5

Total Recalls

2013 - 2016

Year Range

Recent Recalls

2014 HORTON AMBULANCE 503

351 vehicles affected

16V625000

EQUIPMENT:APPLIANCE:AIR CONDITIONER

Halcore Group, Inc. (Halcore) is recalling certain model year 2013-2016 Horton ambulances, models 453, 457, 503, 553, 603, and 623, manufactured November 29, 2013, to March 1, 2016. The affected vehicles may have an improperly sealed Cool-Tech air-conditioning condenser controller wire harness that can allow water to enter the controller housing.

Risk: If water gets into the controller housing, the controller may short circuit, increasing the risk of an electrical fire.

2015 HORTON AMBULANCE 503

18 vehicles affected

15V828000

ELECTRICAL SYSTEM

Halcore Group, Inc. (Halcore) is recalling certain model year 2015 Horton 623, 603, 503, 553 and 453 ambulances manufactured October 1, 2015, to November 16, 2015 and equipped with the i4G system for controlling the ambulance body electrics. Due to a circuit board component being installed incorrectly, the ambulance body may have electrical system failure of the interior and exterior lights, as well as other life saving electrical equipment such as the oxygen system.

Risk: Failure of ambulance body electrics can hinder emergency response and/or rescue efforts, increasing the risk of injury to ambulance patients.

Recent Complaints